For the measurement of pH the HI98191 comes with a HI72911B titanium body pH electrode with a built in temperature sensor for automatic temperature compensated readings. The titanium body is chemically resistant to aggressive chemicals and is virtually unbreakable.

Calibration can be performed up to five points with seven preprogrammed standard pH buffers and five custom ones to achieve a pH accuracy of ±0.002 with ±0.001 pH resolution.

Hanna's unique CAL Check™ feature alerts the user to potential problems during the pH calibration process. By comparing previous calibration data to the current calibration the meter will alert the user with display prompts when the probe needs to be cleaned or is broken and if the pH buffer might be contaminated. After calibration the overall probe condition is displayed as a percentage from 0 to 100% in increments of 10%. The probe condition is affected by both the offset and slope characteristics of the pH electrode and is displayed on the measurement screen on the day of calibration and in the GLP data.

The HI98191 has 15 different ISE sensors pre-programmed with the ion charge and molar weight. By selecting the appropriate ISE in the ISE Setup Menu, the appropriate ion charge and ion molar weight are applied to display ISE slope percentage. Users can then calibrate up to five points with the choice of seven standard values. This meter allows an extensive choice of measurement units (ppm, ppt, g/L, ppb, μg/L, mg/mL, M, mol/L, mmol/L, % w/v, user defined) and has an expanded measuring range of 1.00 x 10-7 to 9.99 x 1010.

Pressing the "AutoHold" virtual key seen in measurement mode, the meter will freeze a stable reading allowing for the reading to be documented or logged. An “out of calibration range” warning can be enabled that will alert the user when a reading is not within the bracket of calibrated pH values.

The log-on-demand mode allows the user to record and save up to 300 samples (100 each for pH, mV, and ISE). This data along with the associated GLP data can then be transferred to a PC with Hanna’s HI920015 micro USB cable and HI92000 software.

Access help at any time at a press of a dedicated button and view content specific information based on the screen that is currently being viewed.

The high contrast graphic LCD is easy to read outdoors in bright sunlight as well in low-lit areas with the backlight. A combination of dedicated and virtual keys allows easy, intuitive operation in a choice of languages.

The compact, durable HI720191 carry case is thermoformed to hold all necessary components for taking a field measurement, including the meter, electrode and beakers.

Specifications

pH Range -2.0 to 20.0 pH; -2.00 to 20.00 pH; -2.000 to 20.000 pH
pH Resolution 0.1 pH; 0.01 pH; 0.001 pH
pH Accuracy ±0.1; ±0.002 pH
pH Calibration up to five point calibration, seven standard buffers available (1.68, 4.01, 6.86, 7.01, 9.18, 10.01, 12.45) + five custom buffers
mV Range ±2000 mV
mV Resolution 0.1 mV
mV Accuracy ±0.2 mV
ISE Range from 1.00 E⁻⁷ to 9.99 Eⁱ⁰ concentration
ISE Resolution 3 digits 0.01; 0.1; 1; 10 concentration
ISE Accuracy ±0.5% of reading (monovalent ions), ±1% of reading (divalent ions)
ISE Calibration up to five point calibration, seven standard solutions (choice of concentration)
Temperature Range -20.0 to 120.0°C (-4.0 to 248.0°F)
Temperature Resolution 0.1°C (0.1°F)
Temperature Accuracy ±0.4°C (±0.8°F) (excluding probe error)
Temperature Compensation manual or automatic from -20.0 to 120.0°C (-4.0 to 248.0°F)
Slope Calibration from 80 to 110%
Electrode/Probe HI72911B Titanium body, pH electrode with internal temperature sensor, BNC connector and 1 m (3.3’ cable)
Input Impedance 10¹² Ohms
Log-on-demand 300 samples (100 each range pH, mV and ISE)
PC Connectivity opto-isolated USB with optional HI 92000 software and micro USB cable
Battery Type/Life 1.5V AA batteries (4) / approximately 200 hours of continuous use without backlight (50 hours with backlight)
Auto-off user selectable: 5, 10, 30, 60 min or can be disabled
Environment 0 to 50°C (32 to 122°F); RH 100% (IP67)
Dimensions 185 x 93 x 35.2 mm (7.3 x 3.6 x 1.4”)
Weight 400 g (14.2 oz.)

 